Coronary Aneurysm Closure With Pericardial Covered Stent Evaluated With Optical Coherence Tomography

Carlos Ortiz a,, Ana Serrador a, José Alberto San Román a

a Servicio de Cardiología, Hospital Clínico Universitario de Valladolid, Instituto de Ciencias del Corazón (ICICOR), Valladolid, Spain

Article

A 62-year-old man with personal history of arterial hypertension and smoking was admitted to our hospital with ST elevation myocardial infarction. Intravenous fibrinolytic therapy was given by emergency services during hospital transfer, which was effective. Routine coronary angiography showed moderate lesion on the proximal left anterior descending, proximal coronary aneurysm (Figure 1, asterisk) and significant lesions at mid segment of circumflex (white arrow) and proximal segment of the second marginal (arrowhead). The case was discussed with the heart team and the decision was made to perform percutaneous coronary intervention.

Figure 1.

Balloon predilation was performed and drug-eluting stent was implanted in second marginal followed by direct drug-eluting stent implantation at mid segment of circumflex, fitting the proximal stent to distal origin of aneurysm. A 3.5 × 18 mm, pericardial, covered stent (Aneugraft, ITGI Medical Ltd., Israel) was implanted at circumflex aneurysm, overlapped to the distal drug-eluting stent with good final result (Figure 1). Several optical coherence tomography pullbacks were conducted in order to confirm correct implantation of pericardial covered stent (Figure 2, asterisk) and correct overlapping with distal drug-eluting stent (Figure 2, white arrow). Patient was discharged on dual antiplatelet therapy with aspirin and clopidogrel for 1 year.

Figure 2.

Although initially designed for vein graft angioplasty, pericardial covered stents have been used to treat other problems in interventional cardiology, such as coronary aneurysm, with limited experience. We present here what is, to the best of our knowledge, the first case to use optical coherence tomography to evaluate the final result of pericardial covered stenting in coronary aneurysm.

Corresponding author: ortiz.bautista.carlos@gmail.com

1885-5857/© 2014 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved

Cookies
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.
Cookies policy
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.