Epsilon Wave in the 12-Lead Electrocardiogram: Is Its Frequency Underestimated?

Javier García-Niebla a,, Adrián Baranchuk b, Antonio Bayés de Luna c

a Centro de Salud Valle del Golfo, Servicios Sanitarios del Área de Salud de El Hierro, Frontera-El Hierro, Sta. Cruz de Tenerife, Spain
b Heart Rhythm Service, Kingston General Hospital, Queen's University, Kingston, Ontario, Canada
c Institut Catala de Ciencies Cardiovasculars, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain

Article

An epsilon wave on a 12-lead electrocardiogram is an important diagnostic criterion for arrhythmogenic right ventricular dysplasia. This delayed right ventricular depolarization manifests as low amplitude electrical potentials between the end of the QRS complex and the beginning of the T wave, most prominently in leads V1 to V3. The epsilon wave is detected in 30% of individuals with arrhythmogenic right ventricular dysplasia; however, the epsilon wave can be masked by excessive low-pass filtering, and the true rate may therefore be higher.

The Figure shows how low-pass filter cutoff frequency influences the detection of the epsilon wave in arrhythmogenic right ventricular dysplasia: At the recommended 150 Hz cutoff frequency (Figure A), the epsilon wave is detected in leads V1-V3. At a 100 Hz cutoff frequency (Figure B), the epsilon wave is attenuated in V1-V2 and absent in V3. At 40 Hz (Figure C), the epsilon wave disappears from leads V1-V3.

Figure.

The low-pass filter cutoff frequency needs to be optimally adjusted to ensure that important clinical information is not lost (the QRS complex and its notches, the J wave and the pacemaker spike). Current electrocardiography guidelines set the cutoff at 150 Hz for adolescents and adults. Nonetheless, in routine clinical practice a 40 Hz cutoff frequency is used to reduce muscle noise and improve the appearance of the trace.

This practice results in the loss of important information, leading us to conclude that the presence of an epsilon wave in patients with arrhythmogenic right ventricular dysplasia may be seriously underestimated.

Corresponding author: jniebla72@hotmail.com

1885-5857/© 2016 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved

Cookies
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.
Cookies policy
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.