Rev Esp Cardiol. 2016;69:346-9 - Vol. 69 Num.03 DOI: 10.1016/j.rec.2015.11.020

Implantation of the Micra Transcatheter Pacing System: Initial Experience in a Single Spanish Center

Marta Pachón a,, Alberto Puchol a, Finn Akerström a, Luis Rodríguez-Padial a, Miguel A. Arias a

a Unidad de Arritmias y Electrofisiología Cardiaca, Servicio de Cardiología, Hospital Virgen de la Salud, Toledo, Spain

Refers to

About the Micra Transcatheter Pacing System
Ángel Morales Martínez de Tejada, Javier Elduayen Gragera
Rev Esp Cardiol. 2016;69:715
Full text - PDF

Article

To the Editor,

The number of pacemakers implanted continues to increase as a result of increased life expectancy. According to the 2013 Spanish Pacemaker Registry,1 the number of first implants per million inhabitants was 567. Despite improved technology and more extensive experience, complications related to pacemaker implantation are still significant, particularly in relation to transvenous leads and the pacemaker pocket.2

Single-component leadless pacemaker systems have been developed to reduce complications. In these systems, the pacemaker and lead are integrated in a single unit, thereby eliminating the risks associated with the leads, pacemaker pocket, and connections.3

In the present letter, we report our initial experience in a single center with implantation of the Micra transcatheter pacemaker system (Medtronic Ibérica, S.A.), approved for clinical use in Europe in June 2015. The device consists of an 8-cm3 capsule, measuring 25.9 × 6.7 mm, designed to provide bipolar VVI-mode cardiac pacing of the right ventricle, with an estimated longevity of 9.6 years. The device also has automated pacing capture threshold management and can be used in magnetic resonance imaging studies without any body restrictions in 1.5 and 3 Tesla systems. The pacemaker is affixed to the ventricular endocardium by 4 self-expanding electrically inert nitinol tines. The device is positioned with the pacemaker and unexpanded tines at the distal end of a steerable catheter, which lies inside an introducer (23-F internal diameter and 27-F external diameter), inserted via the femoral vein. Before the pacemaker is definitively released from the catheter, and once it is positioned in the ventricle, the operator must confirm that the electrical parameters are appropriate and that the system is well affixed to the endocardium by a fluoroscopically guided pull-and-hold test. If the device is not properly affixed or the electrical parameters are deficient, the pacemaker can be removed and the implantation procedure repeated.

The Micra pacemaker was implanted in 10 patients with a standard indication for a permanent pacemaker and with a clinical profile and indication appropriate for VVI pacing. In this initial phase, patients without their own cardiac rhythm were excluded. Prior to implantation, the patients and family members were informed of the characteristics of the new system and the more limited clinical experience with respect to conventional systems. Informed consent was obtained. The general characteristics of the patients and basic data on the procedure are summarized in the Table. The mean age of the patients was 77.1 ± 5.1 years, 6 were men and 4 were women, and 8 had permanent atrial fibrillation as the basal rhythm and 2 had sinus rhythm. The Micra pacemaker was successfully implanted in 10 patients (Figure) by the same operator (M. P.), who had extensive experience in placement of conventional implantable cardiac devices and in electrophysiological procedures with femoral venous and arterial approaches. The first fixation attempt was definitive in 7 patients. In the remaining 3, repositioning was required, although in all patients only 1 further attempt was needed due to deficient electrical parameters after the initial implantation. The mean values of the R wave and impedance were 12.7 ± 4.8 mV and 739 ± 161 ohms, respectively. The mean duration of the implantation procedure was 44.6 ± 7.5 minutes, while fluoroscopy lasted 10.03 ± 2.5 minutes. No patients experienced sustained arrhythmias or blocks during implantation. The femoral puncture site was closed with a Figure 8 suture, with no incidents of note (2 of the patients continued anticoagulant treatment with acenocoumarol during implantation). The clinical and electrical data are summarized in the Table. Of note is the absence of complications requiring intervention and the low rate of repositioning even though these were the first implantations of the new device. At follow-up (mean duration, 55 ± 33 days, range 27-112 days), thresholds (all = 1 V at a pulse width of 0.24 ms), R wave (13.4 ± 5.1 mV), and pacing impedance (633 ± 138 ohms) were very stable.

Patient Characteristics and Implantation and Follow-up Data

  Patient 1 Patient 2 Patient 3 Patient 4 Patient 5 Patient 6 Patient 7 Patient 8 Patient 9 Patient 10
Sex Male Female Male Female Male Male Male Male Female Female
Age, y 70 78 69 80 75 80 81 86 77 75
Weight, kg 88 70 106 46 98 90 85 82 76 59
Height, cm 160 155 174 155 178 173 181 172 162 164
Concomitant conditions Hypertension
Dyslipidemia
Exsmoker
Diabetes mellitus
Hypertension
Dyslipidemia
Diabetes mellitus
Hypertension
Dyslipidemia
Exsmoker
No Hypertension
Dyslipidemia

Hypertension
No Hypertension
Obstructive sleep apnea
Hypertension
Chronic renal failure
COPD
Hypertension
Chronic renal failure
Indication Intermittent syncope due to AF Slow symptomatic AF Intermittent AF Erratic control of AF Bradycardia-tachycardia syndrome Bradycardia-tachycardia syndrome Slow AF Permanent 1st degree AV block and paroxysmal 3rd degree block Erratic control of AF Slow AF
Basal rhythm and frequency, bpm AF AF, 48 AF, 54 AF, 85 SR, 55 AF, 80 AF, 40 SR, 80 AF, 95 AF, 60
QRS, ms 100 90 95 140, left bundle branch block 100 110 95 110, left anterior hemiblock 100 110
LVEF, % 50 60 55 65 60 70 71 60 61 65
Heart disease Ischemic Severe mitral and aortic stenosis; refused surgery Severe mitral regurgitation due to mitral prolapse, with surgical repair Valvular: mechanical mitral prosthesis and tricuspid annuloplasty Moderate degenerative AoVD Hypertensive, mild mitral regurgitation Mild degenerative AoVD Mild degenerative AoVD Moderate AoVD, moderate mitral regurgitation Mechanical mitral prosthesis due to rheumatic disease
Tricuspid value regurgitation and systolic pulmonary pressure Grade II, 40 mmHg Grade II, 50 mmHg Grade III, 49 mmHg Grade II, 43 mmHg Grade III, 25 mmHg Grade II, 35 mmHg Grade I, 32 mmHg Grade I, 30 mmHg Grade II, 37 mmHg Grade II, 50 mmHg
Medications Acenocoumarol
ARB
ASA
Carvedilol
Statins
Amlodipine
Acenocoumarol
ARB
Insulin
Statin
Furosemide
Acenocoumarol
ACEI
Bronchodilators
Statin
Acenocoumarol
Furosemide
Diltiazem
Bisoprolol
Spirolactone
Dabigatran
Nebivolol
ARB
Furosemide
ARB
Hydrochlorothiazide
ASA
Acenocoumarol ACEI Acenocoumarol
ARB
Nitrates
Furosemide
Atenolol
Acenocoumarol
Digoxin
Furosemide
ARB
Particular considerations No No No Previous implantation of transvenous pacemaker, explanted after 1 month due to complicated hematoma No No No No
No
Temporary epicardial pacemaker after cardiac surgery
Total implantation time 51 55 57 43 36 38 37 45 45 39
Fluoroscopy time 15 10.7 13.4 9.5 7.5 8 8.1 10.5 10.3 7.3
Anticoagulation during implantation No No No Yes, INR 2.8 No No No No No Yes, INR 2.2
Anesthesia Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic Conscious sedation and analgesia and local anesthetic
No. of repositioning procedures and reasons 1, threshold > 1.5 V 0 1, threshold > 1.5 V 0 0 0 0 1, threshold > 1.5 V 0 0
Final position Apical Midseptal Midseptal Apical Apical Midseptal Apical Midseptal Apical Midseptal
Implantation
R wave, mV 18.2 14.2 10.4 11.5 4.7 12.5 6.2 16.4 20 12.8
Impendence, ohms 920 520 910 610 660 560 700 990 720 800
Threshold in V at 0.24 ms 1.25 1.25 0.5 0.25 0.5 0.25 0.25 0.5 0.63 0.25
Complications No No No No No No No No No No
Programming VVIR 50-110 bpm VVIR 60-120 bpm VVIR 60-140 bpm VVIR 55-120 bpm VVI 50 bpm VVI 50 bpm VVIR 60-130 bpm VVI 50 bpm VVI 50 bpm VVI 60 bpm
Follow-up, d 112 112 62 62 40 40 33 33 27 27
Follow-up
R wave, mV 20 13 11 12.5 5.7 20 6.4 13.7 19.8 15.1
Impendence, ohms 470 500 880 750 500 580 670 530 600 700
Threshold at 0.24 ms 1 0.88 0.88 0.38 0.38 0.38 0.38 0.5 0.38 0.88
Battery voltage and estimated longevity 3.14, > 8 y 3.14, > 8 y 3.15 V, > 8 y 3.16 V, > 8 y 3.17 V, > 8 y 3.16 V, > 8 y 3.15 V, > 8 y 3.16 V, > 8 y 3.15 V, > 8 y 3.16 V, > 8 y
Complications No No No No No No No No No No
% pacing 8 60 25.5 27.5 3.3 7.5 90.6 21.3 13 25

ACEI, angiotensin converter enzyme inhibitor; AF, atrial fibrillation; AoVD, aortic valve disease; ARB, angiotensin receptor antagonists; ASA, acetylsalicylic acid; bpm, beats per minute; COPD, chronic obstructive pulmonary disease; LVEF, left ventricular ejection fraction; SR, sinus rhythm; V, volts.

A: Image of completely hidden (left) and partially released (right) device in the steerable implantation catheter. B: Radiological detail of the implanted pacemaker in patient 4. C: Posteroanterior chest X-ray in 2 patients with midseptal position and 2 patients with apical position.

Figure. A: Image of completely hidden (left) and partially released (right) device in the steerable implantation catheter. B: Radiological detail of the implanted pacemaker in patient 4. C: Posteroanterior chest X-ray in 2 patients with midseptal position and 2 patients with apical position.

The initial results in terms of electrical performance and safety of the implantation procedure for the Micra transcatheter pacemaker are promising in the series presented here. However, to date there are no studies in humans comparing this new system with conventional pacemaker systems. The largest study reported is the Micra Transcatheter Pacing Study. This was a prospective, multicenter, noncomparative study in which the Micra system was successfully implanted in 719 of 725 attempts (99.2%).4 Major complications were reported in 25 patients, with no cases of systemic infection or dislodgement of the pacemaker. The electrical performance at 6 months of follow-up, assessed in 297 patients, was very satisfactory, with a mean pacing capture threshold of 0.54 V at 0.24 ms, an R-wave of 15.3 mV, and an estimated longevity of 12.5 years (94% of devices had a projected longevity of more than 10 years). This performance was similar to that of modern conventional pacemakers.

Future studies will provide information on the long-term safety and efficacy of this new permanent pacing system. These studies should also aim to confirm the potential benefits of this device compared with conventional lead-based devices.

CONFLICTS OF INTEREST

M. Pachón is a proctor for the Micra device. M.A. Arias is member of the editorial board of Revista Española de Cardiología.

Corresponding author: mpachon@secardiologia.es

Bibliography

1. Coma Samartín R, Cano Pérez O, Pombo Jiménez M. Spanish Pacemaker Registry. Eleventh official report of the Spanish Society of Cardiology Working Group on Cardiac Pacing (2013). Rev Esp Cardiol. 2014;67:1024-38.
2. Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35:1186-94.
3. Miller MA, Neuzil P, Dukkipati SR, Reddy VY. Leadless cardiac pacemakers: back to the future. J Am Coll Cardiol. 2015;66:1179-89.
4. Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2015.

1885-5857/© 2016 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved

Cookies
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.
Cookies policy
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.