Rev Esp Cardiol. 2017;70:774 - Vol. 70 Num.09 DOI: 10.1016/j.rec.2017.01.019

Response to ECG, August 2017

Javier Higueras a,, Julián Palacios-Rubio a, Carmen Olmos a

a Instituto Cardiovascular, Hospital Clínico San Carlos, Universidad Complutense, Madrid, Spain

Refers to

Free articleECG, August 2017
Javier Higueras, Julián Palacios-Rubio, Carmen Olmos
Rev Esp Cardiol. 2017;70:667
Full text - PDF
See the proposed resolutions

Article

A managed ventricular pacing (MVP) algorithm was applied, with the pacemaker working correctly (correct answer, 4). In the upper trace of the Figure, sequential pacing can be observed with an atrioventricular interval of 150 ms, except for the last beat, in which the pacemaker switches to AAI mode. The MPV algorithm periodically makes such checks.1 In the lower trace, the pacemaker is working in AAI mode with atrial pacing (asterisks) when the heart rate drops below the threshold (60 bpm). An atrial pace is present after the sixth QRS: the MVP algorithm will ensure that there is no atrioventricular conduction, but in the next complex (double arrow), the pacemaker emits a safety pace, which with this model has an atrioventricular interval of 80 ms. This is not ventricular capture failure (answers 1 and 2, incorrect) because there is no ventricular pace spike and at no point is failure of atrial capture observed (answer 3, incorrect).

Figure.

Corresponding author: javierhnb@yahoo.es

Bibliography

1. Medtronic Manual Library. Instructions for use and products manuals for Medtronic Adapta/Versa/Sensia Pacemakers. Minneapolis: Medtronic;2009. Available at: http://manuals.medtronic.com/.