Disappearance of Pacemaker Pulse Genarator
Desaparición del generador de marcapasos
Visits
5655
José Luis Martínez Sandea, Antonio Amaro Cendóna, José María García Acuñaa
a Servicio de Cardiología y Unidad Coronaria, Hospital Clínico Universitario, Santiago de Compostela, Spain.
This item has received
5655
Visits
Article information
Full Text
Download PDF
Statistics
Tables (2)
Fig. 1.
Fig. 2.
Show moreShow less
Full Text

An 83-year-old woman with a history of VVI pacemaker implantation 10 years earlier due to complete atrioventricular block. Four months previously at another hospital she had undergone elective replacement of the generator as it was worn out, and she was admitted at the present time for syncope. She lived alone and presented with senile dementia and blindness due to glaucoma in the right eye.

On ECG complete atrioventricular block was seen with a ventricular escape rhythm of 35 beats/minute and morphology of an advanced block of the right branch of the bundle of His. Stimulation flutter was not seen.

On physical examination, the generator pocket appeared edematous and swollen, with continuity solution on the skin, but the generator was not located on palpation.

Chest radiography showed the absence of the pacemaker generator and interruption of the electrode. In addition, continuity fluid was seen on the electrode (Figure 1, arrow).

Fig. 1.

On questioning the patient again, she remembered that she had cut the pacemaker cable and given the generator to the attending cardiologist.

Figure 2 shows the pacemaker generator with intentional cutting of the cable, the partially closed exterior wound (arrow), and the surgical wound, with staples, used for the revision and electrical isolation of the remaining electrode.

Fig. 2.

Staphylococcus aureus was isolated in a wound culture, and treatment was begun with intravenous vancomycin. We decided to proceed to the isolation of the pacemaker electrode and the implantation of a new stimulation system via dissection of the left cephalic vein.

Idiomas
Revista Española de Cardiología (English Edition)

Subscribe to our newsletter

Article options
Tools
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

es en
Política de cookies Cookies policy
Utilizamos cookies propias y de terceros para mejorar nuestros servicios y mostrarle publicidad relacionada con sus preferencias mediante el análisis de sus hábitos de navegación. Si continua navegando, consideramos que acepta su uso. Puede cambiar la configuración u obtener más información aquí. To improve our services and products, we use "cookies" (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.