Publique en esta revista
Información de la revista
Vol. 71. Núm. 11.
Páginas 895-901 (Noviembre 2018)
Compartir
Compartir
Descargar PDF
Más opciones de artículo
ePub
Visitas
115
Vol. 71. Núm. 11.
Páginas 895-901 (Noviembre 2018)
Artículo original
DOI: 10.1016/j.recesp.2017.08.020
Resultados agudos y a largo plazo tras implante contemporáneo de desfibrilador subcutáneo: experiencia en un centro
Acute and Long-term Results After Contemporary Subcutaneous Implantable Cardioverter-defibrillator Implantation: A Single-center Experience
Visitas
115
Miguel A. Arias
Autor para correspondencia
maapalomares@secardiologia.es

Autor para correspondencia: Unidad de Arritmias y Electrofisiología Cardiaca, Hospital Virgen de la Salud, Avda. Barber 30, Planta Semisótano, 45004 Toledo, España.
, Marta Pachón, Finn Akerström, Alberto Puchol, Cristina Martín-Sierra, Luis Rodríguez-Padial
Unidad de Arritmias y Electrofisiología Cardiaca, Servicio de Cardiología, Hospital Virgen de la Salud, Toledo, España
Contenido relaccionado
Rev Esp Cardiol. 2018;71:886-710.1016/j.recesp.2018.04.039
Lucas V.A. Boersma
Este artículo ha recibido
115
Visitas
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (4)
Mostrar másMostrar menos
Tablas (2)
Tabla 1. Características generales de los pacientes
Tabla 2. Datos del implante y prueba de inducción
Mostrar másMostrar menos
Resumen
Introducción y objetivos

El desfibrilador subcutáneo (S-ICD) surge como alternativa al transvenoso. La incidencia de complicaciones es similar, y los choques inapropiados (CI) son más frecuentes que lo observado con programaciones contemporáneas en los transvenosos. Tras aprobarse en 2009 en Europa, se han implementado diversas mejoras. Se expone el resultado en un centro con el implante de S-ICD, cuya experiencia se inició tardíamente, a finales de 2013.

Métodos

Estudio prospectivo observacional con inclusión de pacientes consecutivos con indicación de desfibrilador y sin indicación de estimulación permanente o resincronización cardiaca, a los que se implantó un S-ICD. Se analizaron datos del implante y seguimiento a largo plazo.

Resultados

Se implantó un S-ICD a 50 pacientes que habían superado el cribado electrocardiográfico pertinente. La media de edad era 46,9±15 (15-78) años, y el 72% eran varones. El 38% presentaba una fracción de eyección del ventrículo izquierdo ≤ 35%, y la cardiopatía isquémica fue la más frecuente (34%), seguida de la miocardiopatía hipertrófica (18%). Se usó la técnica intermuscular, con 3 incisiones en el 10% y 2 en el 90%. Se indujo fibrilación ventricular a 49 pacientes, con eficacia del 100% en su conversión. Tras un seguimiento medio de 18,1 (2,3-44,8) meses, no se produjeron complicaciones tardías que requirieran revisión quirúrgica ni CI (0%), y 1 paciente (2%) recibió choques apropiados.

Conclusiones

Las mejoras tecnológicas, de implante y programación, junto con una selección adecuada de pacientes, han permitido obtener unos excelentes resultados agudos y a largo plazo, especialmente por la ausencia de CI y complicaciones que requirieran revisión quirúrgica.

Palabras clave:
Desfibrilador subcutáneo
Muerte súbita
Arritmias ventriculares
Abreviaturas:
CI
DAI
S-ICD
Abstract
Introduction and objectives

The subcutaneous implantable cardioverter-defibrillator (S-ICD) has emerged as an alternative to the transvenous defibrillator. The incidence of complications is similar, with inappropriate shocks (IS) being more frequent than those occurring with contemporary programming of transvenous defibrillators. Several improvements have been implemented after the S-ICD was approved for use in Europe in 2009. This study reports the results of S-ICD use in a single center, whose experience began late, at the end of 2013.

Methods

Prospective observational study including consecutive patients with defibrillator indication and no indication for either permanent pacing or cardiac resynchronization who underwent S-ICD implantation. Implant data and long-term follow-up were analyzed.

Results

An S-ICD was implanted in 50 patients who were deemed suitable after electrocardiographic screening. The mean age was 46.9±15 (range, 15-78) years and 72% were male. Thirty eight percent had left ventricular ejection fraction ≤ 35%. The most frequent heart disease was ischemic heart disease (34%), followed by hypertrophic cardiomyopathy (18%). The intermuscular technique was used, with 3 incisions in 10% and 2 incisions in the remaining 90%. Ventricular fibrillation was induced in 49 patients, with 100% effectiveness in their conversion. After a mean follow-up of 18.1 (range, 2.3-44.8) months, there were no late complications requiring surgical revision, the rate of IS was 0%, and 1 patient (2%) experienced appropriate shocks.

Conclusions

Improvements in technology, implant technique and device programming, along with appropriate patient selection, have led to outstanding acute and long-term results, especially regarding the absence of both IS and complications requiring surgical revision.

Full English text available from:www.revespcardiol.org/en

Keywords:
Subcutaneous defibrillator
Cardiac arrest
Ventricular arrhythmias

Artículo

Opciones para acceder a los textos completos de la publicación
Revista Española de Cardiología
Socio
Socios SEC
Use datos de acceso a SEC en el menú Acceder
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Revista Española de Cardiología

Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9 h a 18 h (GMT+1) excepto los meses de julio y agosto, en los que el horario es de 9 h a 15 h.
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
Email
Idiomas
Revista Española de Cardiología

Suscríbase a la newsletter

Opciones de artículo
Herramientas
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

es en
Política de cookies Cookies policy
Utilizamos cookies propias y de terceros para mejorar nuestros servicios y mostrarle publicidad relacionada con sus preferencias mediante el análisis de sus hábitos de navegación. Si continua navegando, consideramos que acepta su uso. Puede cambiar la configuración u obtener más información aquí. To improve our services and products, we use "cookies" (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.