Original article
Full text access
Conduction Abnormalities and Pacemaker Implantations After SAPIEN 3 Vs SAPIEN XT Prosthesis Aortic Valve Implantation
Trastornos de la conducción e implante de marcapasos tras implante de válvula aórtica SAPIEN 3 comparada con la SAPIEN XT
Visits
9938
Oliver Hussera,
Corresponding author
oliver.husser@gmail.com

Corresponding author: Klinik für Herz-und Kreislauferkrankungen, Deutsches Herzzentrum München, Technische Universität München, Lazarettstrasse 36, 80636 Múnich, Germany.
, Thorsten Kesslera, Christof Burgdorfa, Christian Templinb, Costanza Pellegrinia, Simon Schneiderc, Albert Markus Kasela, Adnan Kastratia,d, Heribert Schunkerta,d, Christian Hengstenberga,d
a Klinik für Herz-und Kreislauferkrankungen, Deutsches Herzzentrum München, Technische Universität München, Munich, Germany
b Universitäres Herzzentrum, Universitätsspital Zürich, Zurich, Switzerland
c Medizinische Klinik und Poliklinik I, Klinikum rechts der Isar, Technische Universität München, Munich, Germany
d Deutsches Zentrum für Herz-Kreislauf-Forschung e.V., Partner Site Munich Heart Alliance, Munich, Germany
Related content
Ramón Rodríguez-Olivares, Freek Nijhoff, Peter R. Stella
This item has received
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (3)
Show moreShow less
Tables (5)
Table 1. Baseline Characteristics of the SAPIEN 3 and SAPIEN XT Subgroups
Table 2. Procedural Data of the SAPIEN 3 and SAPIEN XT Cohorts
Table 3. In-hospital Events of the SAPIEN 3 and SAPIEN XT Subgroups
Table 4. Incidence and Reasons for Permanent Pacemaker Implantations After Transcatheter Aortic Valve Implantation
Table 5. New-onset Intraventricular Conduction Abnormalities According to SAPIEN 3 and SAPIEN XT at Discharge
Show moreShow less
Additional material (1)
Abstract
Introduction and objectives

Transcatheter aortic valve implantation is increasingly used in patients with aortic stenosis. Post-procedural intraventricular conduction abnormalities and permanent pacemaker implantations remain a serious concern. Recently, the Edwards SAPIEN 3 prosthesis has replaced the SAPIEN XT. We sought to determine the incidences of new-onset intraventricular conduction abnormalities and permanent pacemaker implantations by comparing the 2 devices.

Methods

We analyzed the last consecutive 103 patients undergoing transcatheter aortic valve implantation with SAPIEN XT before SAPIEN 3 was used in the next 105 patients. To analyze permanent pacemaker implantations and new-onset intraventricular conduction abnormalities, patients with these conditions at baseline were excluded. Electrocardiograms were recorded at baseline, after the procedure, and before discharge.

Results

SAPIEN 3 was associated with higher device success (100% vs 92%; P=.005) and less paravalvular leakage (0% vs 7%; P<.001). The incidence of permanent pacemaker implantations was 12.6% (23 of 183) with no difference between the 2 groups (SAPIEN 3: 12.5% [12 of 96] vs SAPIEN XT: 12.6% [11 of 87]; P=.99). SAPIEN 3 was associated with a higher rate of new-onset intraventricular conduction abnormalities (49% vs 27%; P=.007) due to a higher rate of fascicular blocks (17% vs 5%; P=.021). There was no statistically significant difference in transient (29% [20 of 69] vs persistent 19% [12 of 64]; P=.168) left bundle branch blocks (28% [19 of 69] vs 17% [11 of 64]; P=.154) when SAPIEN 3 was compared with SAPIEN XT.

Conclusions

We found a trend toward a higher rate of new-onset intraventricular conduction abnormalities with SAPIEN 3 compared with SAPIEN XT, although this did not result in a higher permanent pacemaker implantation rate.

Keywords:
Transcatheter aortic valve implantation
Pacemaker
Conduction abnormalities
Edwards SAPIEN 3
Edwards SAPIEN XT
Abbreviations:
IVCA
LBBB
Resumen
Introducción y objetivos

El implante percutáneo de válvula aórtica se utiliza de manera creciente en pacientes con estenosis aórtica. Los trastornos de la conducción intraventricular y los implantes de marcapasos permanente tras la intervención siguen siendo motivo de gran preocupación. Recientemente, la válvula protésica Edwards SAPIEN XT ha sido reemplazada por la SAPIEN 3. El objetivo de este estudio es determinar la incidencia de trastornos de la conducción intraventricular de nueva aparición y la necesidad de implantes de marcapasos permanente comparando los dos dispositivos.

Métodos

Se analizó a los últimos 103 pacientes consecutivos a los que se realizó implante percutáneo de válvula aórtica SAPIEN XT antes de que se utilizara la válvula SAPIEN 3 y a los 105 pacientes consecutivos a los que se implantó la válvula SAPIEN 3. Con objeto de analizar los implantes de marcapasos permanentes y los trastornos de la conducción intraventricular de nueva aparición, se excluyó del estudio a los pacientes en los que se daban estas características en situación basal. Se registraron electrocardiogramas en situación basal, después de la intervención y antes del alta.

Resultados

El uso del dispositivo SAPIEN 3 se asoció a un porcentaje superior de éxitos de implante del dispositivo (el 100 frente al 92%; p=0,005) y un porcentaje inferior de fuga paravalvular (0 frente al 7%; p<0,001). La incidencia de implantes de marcapasos permanente fue del 12,6% (23 de 183), sin que se apreciaran diferencias entre los dos grupos (SAPIEN 3, el 12,5% [12 de 96]; SAPIEN XT, el 12,6% [11 de 87]; p=0,99). El uso del dispositivo SAPIEN 3 se asoció a mayor frecuencia de trastornos de la conducción intraventricular de nueva aparición (el 49 frente al 27%; p=0,007) como consecuencia de la mayor frecuencia de bloqueos fasciculares (el 17 frente al 5%; p=0,021). No hubo diferencia estadísticamente significativa por lo que respecta a los bloqueos de rama izquierda transitorios frente a persistentes (el 29% [20 de 69] frente al 19% [12 de 64]; p=0,168) al comparar el SAPIEN 3 con el SAPIEN XT (el 28% [19 de 69] frente al 17% [11 de 64]; p=0,154).

Conclusiones

Se observó una tendencia a mayor frecuencia de trastornos de la conducción intraventricular de nueva aparición con SAPIEN 3 que con SAPIEN XT, si bien esto no conllevó mayor frecuencia de implante de marcapasos permanentes.

Palabras clave:
Implante percutáneo de válvula aórtica
Marcapasos
Trastornos de la conducción
Edwards SAPIEN 3
Edwards SAPIEN XT
Full Text
INTRODUCTION

Transcatheter aortic valve implantation (TAVI) is increasingly used as a therapeutic strategy in high-risk patients with severe symptomatic aortic stenosis and has shown encouraging clinical results in large registries1,2 and randomized trials.3,4 However, post-procedural intraventricular conduction abnormalities (IVCA) and the need for permanent pacemaker implantation remain a serious concern.

New-onset IVCA, especially new-onset left bundle branch block (LBBB), are observed in a considerable proportion of patients after TAVI (between 8% and 30%)5,6 with a balloon-expandable valve7–9 and may negatively affect recovery of left ventricular function and increase the risk for permanent pacemaker implantation.10,11

Furthermore, the occurrence of high-grade atrioventricular conduction abnormalities leading to permanent pacemaker implantation is a frequent complication associated with TAVI, with an incidence ranging between 3% and 11.5% with balloon-expandable valves.3,12

Recently, a new generation balloon expandable prosthesis, Edwards SAPIEN 3 (Edwards Lifesciences; Irvine, CA, United States)13 has been introduced replacing the SAPIEN XT in Germany. The SAPIEN 3 device has a higher metal frame and features an outer skirt surrounding the valve's frame, designed to avoid paravalvular leakage. The incidence of new-onset IVCA and rates of permanent pacemaker implantation with this novel prosthesis are still under investigation. Ideally, the improved hemodynamics of novel devices should not be associated with a higher rate of conduction abnormalities.

This study assessed the incidence of new-onset IVCA and the need for permanent pacemaker implantation with the SAPIEN 3 prosthesis in comparison with the SAPIEN XT valve in a consecutive cohort of patients undergoing TAVI.

METHODSPatient Populations and Transcatheter Aortic Valve Implantation Procedure

We prospectively included 222 patients undergoing TAVI at the Klinik für Herz- und Kreislauferkrankungen, Deutsches Herzzentrum München during a 12-month period (SAPIEN XT (n=103) from July 2nd 2013 until February 11th 2014 and SAPIEN 3 (n=105) from January 31st to June 26th 2014). Patients receiving other valve types were not included (n=14).

To determine the incidence of new permanent pacemaker implantation after TAVI, patients carrying a pacemaker (n=24) as well as 1 patient who underwent emergency TAVI were excluded. Additionally, to analyze the incidence of new-onset IVCA after TAVI, patients with IVCA at baseline were excluded (n=48) (Figure 1).

Figure 1.

Study design and flow of patients. In total, 210 patients were suitable for transfemoral TAVI using an Edwards valve during the prespecified observation interval. Of these, 103 and 105 patients received SAPIEN XT and SAPIEN 3 valves, respectively. Two patients did not undergo TAVI due to inadequate access site (n=1) and short distance to the coronary ostium (n=1). After exclusion of patients with a prior permanent pacemaker, and 1 patient who underwent transcatheter aortic valve implantation in an emergency situation, 183 patients were available for new permanent pacemaker implantation analysis comparing the 2 devices. New-onset IVCA were analyzed in patients with normal electrocardiograms at baseline (n=135). LAHB, left anterior hemiblock; LBBB, left bundle branch block; RBBB, right bundle branch block; uIVCA, unspecific intraventricular conduction abnormality.

(0.31MB).

TAVI was performed in a hybrid operating room using the transfemoral approach in all patients.

Multislice Computed Tomography and Prosthesis Size Selection

Aortic annulus measurements were assessed in multiple plane reconstruction according to the guidelines of the Society of Cardiovascular Computed Tomography.14 The area and perimeter of the virtual aortic annulus were obtained by direct planimetry. The minimal and maximum diameters were determined and the effective diameter was calculated as the sum of the minimum and maximum diameter divided by 2. The eccentricity of the aortic annulus was calculated as the eccentricity index using the formula [1-minimum diameter/maximum diameter]. The closer this index comes to zero, the more circular the aortic annulus. Dedicated Food and Drug Administration approved software (OsiriX MD 3.9.4, Pixmeo, Switzerland) was used. The respective prosthesis size was selected according to the manufacturers’ recommendations but the final decision on prosthesis size was left at the discretion of the physicians performing the procedure.

The SAPIEN XT and SAPIEN 3 heart valves are available in 23-, 26- and 29-mm sizes covering a nominal area of 4.15cm2, 5.31cm2, and 6.21cm2 and 4.06cm2, 5.19cm2, and 6.48cm2, respectively. The percentage of oversizing according to area was calculated using the formula (nominal prosthesis area/patient annular area-1)×100.

Electrocardiogram Analysis and Permanent Pacemaker Implantations

Electrocardiograms were recorded at a minimum of 3 time points: on admission (at least 24hours before TAVI), immediately after TAVI, and before discharge. Electrocardiogram data were reviewed by 2 physicians blinded to clinical data according to current recommendations.15 Controversial cases were solved by consensus. The following electrocardiographic variables were evaluated at all time points: rhythm (sinus rhythm, atrial fibrillation or flutter), heart rate, PQ duration and QRS intervals, atrioventricular conduction disturbances (none, grade I, II and III), and IVCA (complete bundle branch block, including left and right bundle branch block separately), fascicular blocks (including left anterior and left posterior hemiblock and incomplete left and right bundle branch block), and nonspecific IVCA. New-onset IVCA, present after TAVI but not at discharge, were classified as transient, whereas IVCA still present at discharge were termed persistent.

The decision to implant a permanent pacemaker was left at the discretion of the physicians in charge of the patients’ treatment. The indications were recorded in every case. In patients with normal intraventricular conduction at baseline undergoing permanent pacemaker implantation after TAVI, the last available electrocardiogram prior to permanent pacemaker implantation was used for the analysis of new-onset IVCA.

Follow-up and Definition of Outcomes

All baseline, procedural, postoperative and electrocardiogram data were prospectively recorded. In-hospital outcome, device success and procedural complications were categorized using the updated criteria defined by the Valve Academic Research Consortium.16

Statistical Analysis

Continuous variables are expressed as mean±standard deviation or as median [interquartile range] and compared using the paired or unpaired Student t test or Mann-Whitney U test, as appropriate. Discrete variables were compared with the chi-square test or Fisher's exact test, as appropriate. To assess the influence of the learning curve with the novel device, the cohort of patients was divided into quartiles, and the incidence of new permanent pacemaker implantation and new persistent LBBB were analyzed. A 2-sided P-value of<.05 was considered statistically significant. SPSS 20 (IBM Corporation; United States) was used.

RESULTSBaseline Characteristics

Baseline characteristics according to prosthesis type are shown in Table 1. There were no statistically significant differences between the 2 groups except a higher rate of previous percutaneous coronary intervention in the SAPIEN XT cohort (44% vs 27%; P=.019).

Table 1.

Baseline Characteristics of the SAPIEN 3 and SAPIEN XT Subgroups

  SAPIEN 3 (n=96)  SAPIEN XT (n=87)  P 
Clinical features
Age, y  80±81±.279 
Male sex, n (%)  51 (53)  46 (53)  .999 
Logistic EuroSCORE, %  14±15±10  .588 
Diabetes mellitus, n (%)  29 (30)  23 (26)  .624 
Hypertension, n (%)  85 (89)  80 (92)  .468 
Hypercholesterolemia, n (%)  66 (69)  67 (77)  .246 
Smoking, n (%)  19 (20)  17 (20)  .966 
Previous SAVR, n (%)  2 (2)  1 (1)  .999 
Previous coronary artery bypass graft, n (%)  5 (5)  11 (13)  .114 
Previous PCI, n (%)  26 (27)  38 (44)  .019 
Previous malignoma, n (%)  21 (22)  18 (21)  .859 
Previous myocardial infarction, n (%)  5 (5)  8 (9)  .390 
Previous stroke, n (%)  10 (10)  3 (3)  .086 
Glomerular filtration rate, mL/min  53±20  52±21  .621 
Chronic obstructive pulmonary disease, n (%)  13 (14)  10 (12)  .824 
Peripheral arteriopathy, n (%)  12 (13)  5 (6)  .133 
NHYA III/IV, n (%)  60 (63)  55 (63)  .999 
Echocardiography
Left ventricular ejection fraction ≤ 40%, n (%)  18 (19)  14 (16)  .699 
Mean gradient, mmHg  41±17  46±18  .074 
Aortic valve area, cm2  0.75±0.22  0.72±0.20  .355 
Electrocardiogram
Atrial fibrillation, n (%)  26 (27)  26 (28)  .744 
LBBB, n (%)  1 (1)  4 (5)  .193 
RBBB, n (%)  11 (12)  9 (19)  .809 
MSCT measurements of the aortic annulus
Minimal diameter, mm  20.94±2.64  20.83±2.42  .761 
Maximal diameter, mm  26.82±2.78  26.85±2.35  .920 
Effective diameter, mm  23.94±2.60  23.84±2.16  .795 
Perimeter, mm  76.70±7.93  76.79±6.95  .935 
Area, cm2  4.58±0.95  4.52±0.99  .667 
Eccentricity index  0.22±0.07  0.22±0.07  .585 
% Oversizing by area  12 [4-17]  16 [9-25]  .002 

LBBB, left bundle branch block; LVEF, left ventricular ejection fraction; MSCT, multislice computer tomography; NYHA, New York Heart Association; PCI, percutaneous coronary intervention; RBBB, right bundle branch block; SAVR, surgical aortic valve replacement.

Diabetes mellitus: insulin-dependent as well as noninsulin dependent diabetes mellitus. Hypertension: resting blood pressure > 140/90mmHg and/or medication with antihypertensive agents. Hypercholesterolemia: total cholesterol > 200mg/dL and/or medication with cholesterol synthesis inhibitors. Smoking: present and past smokers. Atrial fibrillation denotes rhythm on admission electrocardiogram.

Procedural Data, In-hospital Outcome and Adverse Events

Significantly more procedures were performed under conscious sedation using the SAPIEN 3 device. Use of the SAPIEN 3 valve was associated with a significantly higher device success rate (100% vs 92%; P=.005) and a lower prevalence of higher than grade 1 post-procedural aortic regurgitation (0% vs 7%; P=.009). Procedural time was significantly shorter for patients treated with the SAPIEN 3 prosthesis compared with its predecessor (59 vs 75minutes; P<.001) (Table 2). While aortic annulus measurements on multislice computer tomography did not differ between the 2 groups, oversizing by area was significantly larger in the case of SAPIEN XT compared with SAPIEN 3 (16% vs 12%; P=.002) (Table 1).

Table 2.

Procedural Data of the SAPIEN 3 and SAPIEN XT Cohorts

  SAPIEN 3 (n=96)  SAPIEN XT (n=87)  P 
General anesthesia  56 (58)  70 (81)  .002 
Procedural time, min  59 [50-71]  75 [60-87]  <.001 
Fluoroscopy time, min  13 [10-17]  14 [11-20]  .148 
Contrast, mL  100 [75-140]  105 [84-139]  .809 
Coronary impairment  0 (0)  2 (2)  .225 
Postoperative emergency sternotomy  1 (1)  2 (2)  .605 
Post-procedural aortic regurgitation (> 1)  0 (0)  6 (7)  .009 
Device success  96 (100)  80 (92)  .005 
Procedural success  96 (100)  85 (98)  .225 
23 mm  38 (40)  21 (24)  .029 
26 mm  36 (38)  49 (56)   
29 mm  22 (23)  17 (20)   

Data are expressed as no. (%), or median [interquartile range].

Patients treated with SAPIEN 3 had a significantly shorter stay in the intensive care unit (1 day vs 2 days; P<.001) and in hospital (6 days vs 7 days; P<.001). There was no statistically significant difference in in-hospital events between the 2 groups (Table 3).

Table 3.

In-hospital Events of the SAPIEN 3 and SAPIEN XT Subgroups

  SAPIEN 3 (n=96)  SAPIEN XT (n=87)  P 
Need for acute dialysis  1 (1)  2 (2)  .605 
Major vascular complications  2 (2)  4 (5)  .426 
Life-threatening bleeding  3 (3)  5 (6)  .481 
Myocardial infarction (< 72 h)  0 (0)  2 (2)  .135 
Disabling stroke  2 (2)  2 (2)  .999 
Death  0 (0)  3 (3)  .105 
Days on ICU  1 [1-2]  2 [1-3]  <.001 
Days in hospital  6 [5-7]  7 [6-8]  .001 

ICU, intensive care unit.

Data are expressed as no. (%), or median [interquartile range].

New Permanent Pacemaker Implantations

The incidence of permanent pacemaker implantation after TAVI was 12.5% in the SAPIEN 3 (12 of 96) and 12.6% the SAPIEN XT groups (11 of 87). The baseline characteristics of patients with and without permanent pacemaker implantation after TAVI are depicted in the Table of the supplementary material. The reasons for permanent pacemaker implantation after TAVI are displayed in Table 4. There were no significant differences between the 2 groups regarding relative and absolute indications for permanent pacemaker implantation. Patients requiring permanent pacemaker implantation had less normal conduction (10 of 23 [44%] vs 125 of 160 [78%]; P<.001), more right bundle branch block (8 of 23 [35%] vs 12 of 160 [8%]; P<.001) and a longer QRS interval (113±26ms vs 99±20ms; P=.023) than patients not requiring permanent pacemaker implantation. Multislice computer tomography data and the percentage of oversizing by area did not differ between patients requiring permanent pacemaker implantation and those who did not (Table of the supplementary material).

Table 4.

Incidence and Reasons for Permanent Pacemaker Implantations After Transcatheter Aortic Valve Implantation

  SAPIEN 3 (n=96)  SAPIEN XT (n=87)  P 
Permanent pacemaker implantation (%)  12 (12.5)  11 (12.6)  .999 
Absolute indication   
Persistent AV-block III   
Relative indication  10   
Intermittent AV-block III   
Symptomatic bradycardia   
Intermittent AV-block II   
Symptomatic bradycardia and LBBB   
LBBB and AV-block I   

AV-block, atrioventricular block; LBBB, left bundle branch block.

The incidence and reasons for permanent pacemaker implantations after transcatheter aortic valve implantation in both cohorts was determined after exclusion of patients carrying a pacemaker (n=24) at baseline.

The rate of permanent pacemaker implantation in the SAPIEN 3 cohort was stable across consecutive tertiles, indicating no decrease of permanent pacemaker implantation rate with growing experience using the new device (P for trend .803).

New-onset Intraventricular Conduction Abnormalities

In total, 135 patients without IVCA in the baseline electrocardiogram underwent TAVI, of whom 66 and 69 were treated with the SAPIEN XT and the SAPIEN 3 valve, respectively. Follow-up electrocardiogram data were not available in 2 patients treated with the SAPIEN XT prosthesis due to intraprocedural death. New-onset IVCA before discharge with the 2 devices are depicted in Figure 2 and Table 5. Overall, patients treated with the SAPIEN 3 prosthesis had a significantly lower rate of normal intraventricular conduction (51% vs 73%; P=.007) mainly due to a significantly higher rate of new left anterior hemiblock (17% vs 5%; P=.021). The incidence of new-onset bundle branch block did not differ between the 2 devices.

Figure 2.

New-onset intraventricular conduction abnormalities with SAPIEN 3 compared with SAPIEN XT. Absolute numbers for the respective IVCA are displayed in the graphs. Thirty-five (51%) and 47 (73%) patients in the SAPIEN 3 and SAPIEN XT cohorts did not display new-onset IVCA at discharge (P=.007), respectively (see Table 5 for details). IVCA, intraventricular conduction abnormality; uIVCA, unspecific intraventricular conduction abnormality.

* In 2 patients, no discharge electrocardiographic data were available due to intraprocedural death.

(0.07MB).
Table 5.

New-onset Intraventricular Conduction Abnormalities According to SAPIEN 3 and SAPIEN XT at Discharge

  SAPIEN 3 (n=69)  SAPIEN XT (n=64)  P 
Any IVCA  34 (49)  17 (27)  .007 
Bundle branch block  19 (28)  12 (19)  .231 
LBBB  19  11   
RBBB   
Fascicular Block  12 (17)  3 (5)  .021 
LAHB  12   
LPHB   
uIVCA  3 (4)  2 (3)  .711 

IVCA, intraventricular conduction abnormality; LAHB, left anterior hemiblock; LBBB, left bundle branch block; LPHB, left posterior hemiblock; RBBB, right bundle branch block; IVCA, intraventricular conduction abnormality; uIVCA, unspecific intraventricular conduction abnormalities.

The trend in new-onset IVCA from the post-procedural phase to discharge in the entire population is depicted in Figure 3. New transient LBBB and persistent LBBB occurred in 24% (32 of 133) and 22% (30 of 133) of patients, respectively. There was a trend towards a higher number in the occurrence of transient (29% [20 of 69] vs 19% [12 of 64]; P=.168) and persistent LBBB (28% [19 of 69] vs 17% [11 of 64]; P=.154) in SAPIEN 3 compared with SAPIEN XT patients, although there was no statistically significant difference. Data stratified into SAPIEN 3 and SAPIEN XT treated patients are displayed in the Figure of the supplementary material.

Figure 3.

Trend in new-onset intraventricular conduction abnormalities after transcatheter aortic valve implantation in the entire study population. For details see text. Absolute numbers for the respective IVCA are displayed in the graphs. IVCA, intraventricular conduction abnormality; uIVCA, unspecific intraventricular conduction abnormality.

* In 2 patients, no discharge electrocardiographic data were available due to intraprocedural death.

(0.18MB).

To analyze whether persistent new-onset LBBB was affected by the experience with the new device, we looked at the tertiles of consecutive patients, but observed a stable rate in the SAPIEN 3 cohort with time (P for trend .876).

DISCUSSION

Recently, the Edwards SAPIEN 3 device replaced the SAPIEN XT valve in Europe. The present study reports our initial experience with this novel device and focuses on the incidence of new-onset IVCA and permanent pacemaker implantation in comparison with its predecessor. We found a trend towards a higher rate of new-onset IVCA in the SAPIEN 3 cohort while permanent pacemaker implantation rates were similar compared with SAPIEN XT.

Permanent Pacemaker Implantations After Transcatheter Aortic Valve Implantation

The incidence of permanent pacemaker implantation with the SAPIEN XT valve ranges between 5% and 12%.3,17 In our patients, the permanent pacemaker implantation rate with the SAPIEN XT valve was similar to these previous observations. Our analysis also confirmed that pre-existing right bundle branch block was associated with permanent pacemaker implantation after TAVI.18,19

Pacemaker dependency may limit clinical benefit from TAVI due to lack of atrioventricular-synchrony and right ventricular pacing,20 although some authors did not observe an effect of permanent pacemaker implantation on clinical outcome in terms of survival.12,21 Additionally, permanent pacemaker implantation after TAVI has been identified as an important cause of prolonged hospital stay, thereby increasing procedural costs.22 Thus, the need for permanent pacemaker implantation as a specific aspect of clinical outcome after TAVI is of major importance.

In the present study, we report the permanent pacemaker implantation rate using the novel SAPIEN 3 device. Interestingly, despite a trend toward a higher rate of new-onset IVCA, use of the SAPIEN 3 device did not result in a higher rate of permanent pacemaker implantation during hospitalization compared with its predecessor, SAPIEN XT. As far as pacemaker implantations after TAVI with the SAPIEN 3 is concerned, our data are in line with emerging data indicating a rate of 13% to 30%.23–25

The specific indications for permanent pacemaker implantation after TAVI are frequently not reported in the literature. While there are absolute indications, eg, persistent complete atrioventricular block, several clinical scenarios lead to relative indications. To date, no consensus has been reached for clear permanent pacemaker implantation indications after TAVI. Current recommendations include electrocardiographic monitoring of TAVI patients and acknowledge that there is a proportion of patients at higher risk who may require prolonged monitoring.26 There is also emerging evidence that atrioventricular-conduction abnormalities after TAVI may resolve over time. Indeed, one study reported resolution of perioperative atrioventricular block in more than half of the patients after implantation of a self-expandable valve.27 Indications for permanent pacemaker implantation in the present study were left at the discretion of the physician in charge of the patients’ treatment. We report a considerable number of patients with a relative indication for permanent pacemaker implantation. Due to the elevated risk for events in the high-risk TAVI population our, policy for permanent pacemaker implantation was more liberal.

New-onset Left Bundle Branch Block After Transcatheter Aortic Valve Implantation. Incidence and Clinical Impact

New-onset LBBB is a frequent observation after TAVI with balloon-expandable valves7–9 and is regarded as a direct result of procedure-induced injury to the conduction system. Indeed, most conduction abnormalities already occur during valvuloplasty28 and a tendency to reversal in a small proportion of patients has been reported.6 In the present study, we found that 22% of LBBB occurring after transcatheter aortic valve implantation resolved before discharge. However, 6 patients with no LBBB after TAVI developed LBBB before discharge so the absolute number of complete LBBB at discharge did not differ significantly.

The clinical relevance of LBBB after TAVI remains a subject of ongoing investigations. While no impact on mortality has been observed, there is evidence that new-onset LBBB after transcatheter aortic valve implantation may be associated with a higher risk of late permanent pacemaker implantation and impaired recovery of left ventricular function.6,8,10,11

The incidence of new-onset LBBB after TAVI with the SAPIEN XT valve ranges between 8% and 30%.5,6 Our current study reports the incidence of new-onset IVCA with a nonsignificant trend toward a higher incidence of new-onset persistent LBBB of 28% using the SAPIEN 3 device compared with 17% in the SAPIEN XT group. Additionally, there was a significantly increased rate of overall new-onset IVCA in the SAPIEN 3 group.

The potential underlying mechanisms for these findings may be due to differences in prosthesis design and implantation mechanisms. SAPIEN 3 features a higher metal frame than SAPIEN XT and an outer skirt to minimize paravalvular regurgitation. Differences in valve deployment with more foreshortening of SAPIEN 3 within the left ventricular outflow tract during implantation may lead to increased mechanical trauma to the cardiac conduction system, which may potentially explain the higher incidence of fascicular blocks observed in the present study. Another relevant factor may be the height of implantation within the aortic annulus, which was not analyzed in the present study. Lastly, the appropriateness of sizing algorithms issued by the manufacturer need to be confirmed. Future research should also focus on the identification of the underlying pathophysiological mechanisms.

Strengths and Limitations

The present study is limited by its sample size and is not based on a randomized comparison of the 2 devices. The findings of our study are strengthened by the fact that all procedures were performed by the same team in a highly experienced TAVI center and that rates of IVCA reported with the SAPIEN XT device are similar to previously published data. However, the influence of a learning curve in the SAPIEN 3 group may not be completely ruled out due to previous experience with SAPIEN XT. In addition, although new-onset LBBB was increased by 58% in the SAPIEN 3 cohort compared with our SAPIEN XT cohort, the observed incidence of 28% is still in the range reported previously for the SAPIEN XT valve.5,6 Accordingly, LBBB rates with the SAPIEN 3 in other centers may vary considerably.

The depth of implantation has been identified as a predictor of new conduction abnormalities.6,29,30 Unfortunately, this parameter was not available for all patients in the present study.

Although the in-hospital rate of new permanent pacemaker implantations did not differ, a possible influence of the higher rate of new-onset conduction abnormalities on the incidence of late permanent pacemaker implantation with SAPIEN 3 remains to be determined by larger studies with long-term follow-up.

CONCLUSIONS

In our initial experience, we found a trend toward a higher rate of new-onset IVCA with SAPIEN 3; however, we observed a similar rate of permanent pacemaker implantation compared with its predecessor, the SAPIEN XT prosthesis. Further research focusing on potential underlying mechanisms and possible strategies to reduce the incidence of new-onset IVCA is warranted.

CONFLICTS OF INTEREST

C. Hengstenberg and A.M. Kasel are proctors for Edwards Lifesciences.

References
[1]
C.W. Hamm, H. Möllmann, D. Holzhey, A. Beckmann, C. Veit, H.R. Figulla, GARY-Executive Board, et al.
The German Aortic Valve Registry (GARY): in-hospital outcome.
Eur Heart J., 35 (2014), pp. 1588-1598
[2]
M. Gilard, H. Eltchaninoff, B. Iung, P. Donzeau-Gouge, K. Chevreul, J. Fajadet, FRANCE 2 Investigators, et al.
Registry of transcatheter aortic-valve implantation in high-risk patients.
N Engl J Med., 366 (2012), pp. 1705-1715
[3]
M.B. Leon, C.R. Smith, M. Mack, D.C. Miller, J.W. Moses, L.G. Svensson, PARTNER Trial Investigators, et al.
Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery.
N Engl J Med., 363 (2010), pp. 1597-1607
[4]
C.R. Smith, M.B. Leon, M.J. Mack, D.C. Miller, J.W. Moses, L.G. Svensson, PARTNER Trial Investigators, et al.
Transcatheter vs surgical aortic-valve replacement in high-risk patients.
N Engl J Med., 364 (2011), pp. 2187-2198
[5]
I. Franzoni, A. Latib, F. Maisano, C. Costopoulos, L. Testa, F. Figini, et al.
Comparison of incidence and predictors of left bundle branch block after transcatheter aortic valve implantation using the CoreValve vs the Edwards valve.
Am J Cardiol., 112 (2013), pp. 554-559
[6]
M. Urena, M. Mok, V. Serra, E. Dumont, L. Nombela-Franco, R. DeLarochellière, et al.
Predictive factors and long-term clinical consequences of persistent left bundle branch block following transcatheter aortic valve implantation with a balloon-expandable valve.
J Am Coll Cardiol., 60 (2012), pp. 1743-1752
[7]
P. Houthuizen, R.M. van der Boon, M. Urena, N. van Mieghem, G.B. Brueren, T.T. Poels, et al.
Occurrence, fate and consequences of ventricular conduction abnormalities after transcatheter aortic valve implantation.
EuroIntervention., 9 (2014), pp. 1142-1150
[8]
T.M. Nazif, M.R. Williams, R.T. Hahn, S. Kapadia, V. Babaliaros, J. Rodés-Cabau, et al.
Clinical implications of new-onset left bundle branch block after transcatheter aortic valve replacement: analysis of the PARTNER experience.
Eur Heart J., 35 (2013), pp. 1599-1607
[9]
M. Urena, J.G. Webb, A. Cheema, V. Serra, S. Toggweiler, M. Barbanti, et al.
Impact of new-onset persistent left bundle branch block on late clinical outcomes in patients undergoing transcatheter aortic valve implantation with a balloon-expandable valve.
JACC Cardiovasc Interv., 7 (2014), pp. 128-136
[10]
R. Hoffmann, R. Herpertz, S. Lotfipour, Ö. Aktug, K. Brehmer, W. Lehmacher, et al.
Impact of a new conduction defect after transcatheter aortic valve implantation on left ventricular function.
JACC Cardiovasc Interv., 5 (2012), pp. 1257-1263
[11]
A. Tzikas, B.M. van Dalen, N.M. van Mieghem, J.L. Gutierrez-Chico, R.J. Nuis, F. Kauer, et al.
Frequency of conduction abnormalities after transcatheter aortic valve implantation with the Medtronic-CoreValve and the effect on left ventricular ejection fraction.
Am J Cardiol., 107 (2011), pp. 285-289
[12]
M. Urena, J.G. Webb, C. Tamburino, A.J. Muñoz-García, A. Cheema, A.E. Dager, et al.
Permanent pacemaker implantation following transcatheter aortic valve implantation: impact on late clinical outcomes and left ventricular function.
Circulation., 129 (2013), pp. 1233-1243
[13]
R.K. Binder, J. Rodés-Cabau, D.A. Wood, M. Mok, J. Leipsic, R. De Larochellière, et al.
Transcatheter aortic valve replacement with the SAPIEN 3: a new balloon-expandable transcatheter heart valve.
JACC Cardiovasc Interv., 6 (2013), pp. 293-300
[14]
S. Achenbach, V. Delgado, J. Hausleiter, P. Schoenhagen, J.K. Min, J.A. Leipsic.
SCCT expert consensus document on computed tomography imaging before transcatheter aortic valve implantation (TAVI)/transcatheter aortic valve replacement (TAVR).
J Cardiovasc Comput Tomogr., 6 (2012), pp. 366-380
[15]
B. Surawicz, R. Childers, B.J. Deal, L.S. Gettes, J.J. Bailey, A. Gorgels, et al.
American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; American College of Cardiology Foundation; Heart Rhythm Society. AHA/ACCF/HRS recommendations for the standardization and interpretation of the electrocardiogram: part III: intraventricular conduction disturbances: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society. Endorsed by the International Society for Computerized Electrocardiology.
J Am Coll Cardiol., 53 (2009), pp. 976-981
[16]
A.P. Kappetein, S.J. Head, P. Généreux, N. Piazza, N.M. Van Mieghem, E.H. Blackstone, et al.
Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document.
J Am Coll Cardiol., 60 (2012), pp. 1438-1454
[17]
H. Eltchaninoff, A. Prat, M. Gilard, A. Leguerrier, D. Blanchard, G. Fournial, FRANCE Registry Investigators, et al.
Transcatheter aortic valve implantation: early results of the FRANCE (FRench Aortic National CoreValve and Edwards) registry.
Eur Heart J., 32 (2011), pp. 191-197
[18]
L. Roten, P. Wenaweser, E. Delacrétaz, G. Hellige, S. Stortecky, H. Tanner, et al.
Incidence and predictors of atrioventricular conduction impairment after transcatheter aortic valve implantation.
Am J Cardiol., 106 (2010), pp. 1473-1480
[19]
G.C. Siontis, P. Jüni, T. Pilgrim, S. Stortecky, L. Büllesfeld, B. Meier, et al.
Predictors of permanent pacemaker implantation in patients with severe aortic stenosis undergoing TAVR: a meta-analysis.
J Am Coll Cardiol., 64 (2014), pp. 129-140
[20]
L.F. Tops, M.J. Schalij, J.J. Bax.
The effects of right ventricular apical pacing on ventricular function and dyssynchrony implications for therapy.
J Am Coll Cardiol., 54 (2009), pp. 764-776
[21]
L. Buellesfeld, S. Stortecky, D. Heg, S. Hausen, R. Mueller, P. Wenaweser, et al.
Impact of permanent pacemaker implantation on clinical outcome among patients undergoing transcatheter aortic valve implantation.
J Am Coll Cardiol., 60 (2012), pp. 493-501
[22]
K. Chevreul, M. Brunn, B. Cadier, G. Haour, H. Eltchaninoff, A. Prat, FRANCE registry investigators, et al.
Cost of transcatheter aortic valve implantation and factors associated with higher hospital stay cost in patients of the FRANCE (FRench Aortic National CoreValve and Edwards) registry.
Arch Cardiovasc Dis., 106 (2013), pp. 209-219
[23]
M.I. Murray, N. Geis, S.T. Pleger, K. Kallenbach, H.A. Katus, R. Bekeredjian, et al.
First experience with the new generation edwards sapien 3 aortic bioprosthesis: procedural results and short term outcome.
J Interv Cardiol., 28 (2015), pp. 109-116
[24]
J. Webb, G. Gerosa, T. Lefèvre, J. Leipsic, M. Spence, M. Thomas, et al.
Multicenter evaluation of a next-generation balloon-expandable transcatheter aortic valve.
J Am Coll Cardiol., 64 (2014), pp. 2235-2243
[25]
G. Tarantini, M. Mojoli, P. Purita, M. Napodano, A. D’Onofrio, A. Frigo, et al.
Unravelling the (arte)fact of increased pacemaker rate with the Edwards SAPIEN 3 valve.
EuroIntervention., (2014),
[26]
D.R. Holmes, M.J. Mack.
Transcatheter valve therapy a professional society overview from the american college of cardiology foundation and the society of thoracic surgeons.
J Am Coll Cardiol., 58 (2011), pp. 445-455
[27]
R.M. van der Boon, N.M. van Mieghem, D.A. Theuns, R.J. Nuis, S.T. Nauta, P.W. Serruys, et al.
Pacemaker dependency after transcatheter aortic valve implantation with the self-expanding Medtronic CoreValve System.
Int J Cardiol., 168 (2013), pp. 1269-1273
[28]
R.J. Nuis, N.M. van Mieghem, C.J. Schultz, A. Tzikas, R.M. van der Boon, A.M. Maugenest, et al.
Timing and potential mechanisms of new conduction abnormalities during the implantation of the Medtronic CoreValve System in patients with aortic stenosis.
Eur Heart J., 32 (2011), pp. 2067-2074
[29]
Ö. Aktug, G. Dohmen, K. Brehmer, R. Koos, E. Altiok, V. Deserno, et al.
Incidence and predictors of left bundle branch block after transcatheter aortic valve implantation.
Int J Cardiol., 160 (2012), pp. 26-30
[30]
R.M. van der Boon, P. Houthuizen, M. Urena, T.T. Poels, N.M. van Mieghem, G.R. Brueren, et al.
Trends in the occurrence of new conduction abnormalities after transcatheter aortic valve implantation.
Catheter Cardiovasc Interv., 85 (2015), pp. E144-E152
Copyright © 2015. Sociedad Española de Cardiología
Idiomas
Revista Española de Cardiología (English Edition)

Subscribe to our newsletter

View newsletter history
Article options
Tools
Supplemental materials