INTRODUCTION
Optical coherence tomography (OCT) is a new intracoronary imaging modality that provides in vivo high-resolution images of the coronary artery.1 The OCT imaging technique is, in general, similar to that of intravascular ultrasound (IVUS), but whereas IVUS works by applying radiofrequency radiation, OCT uses light radiation. In both techniques the imaging probe is automatically pulled back through the coronary segment of interest. In contrast to IVUS, in OCT acquisition the coronary vessel needs to be cleared of blood. This is necessary since red blood cells are non-transparent causing multiple light scattering and substantial signal attenuation. This was achieved in the first commercially available OCT system by proximal occlusion of the vessel with a dedicated low-pressure balloon and simultaneous distal flush delivery (occlusive technique) during OCT probe pullback.2 However, the recent increase in pullback speed (from 1 mm/s to 3 mm/s) enabled OCT pullback while displacing blood by continuous flush (x-ray contrast medium) delivery through the guide catheter without the need for vessel occlusion (non-occlusive technique).3 The impact of these 2 different acquisition techniques on lumen dimension when measured by OCT has not yet been reported.
The present study had 2 aims: to compare lumen size when measured by IVUS, OCT and histology in human coronary arteries ex vivo; to compare, in vivo, intracoronary lumen size when measured in patients by OCT with balloon occlusion, OCT without balloon occlusion, and IVUS.
METHODS
Ex Vivo Human Coronary Arteries
Specimens
Ten left anterior descending coronary artery (LAD) segments were excised within 24 h postmortem, 1 cm proximal to the bifurcation with the left circumflex artery (LCX) and fixed in 4% formaldehyde. Standard 6 F coronary sheaths (Arrow, Reading, PA, USA) were introduced and fixed at the distal and proximal ends of the arteries.
Intracoronary Imaging
For intracoronary imaging, the specimens were immersed in a saline bath at room temperature and perfused with saline at physiologic pressure. IVUS data were acquired using the AtlantisTM 40 MHz catheter with an automated motorized pullback device at a constant speed of 0.5 mm/s. OCT data were acquired using a commercially available system for intracoronary imaging and a 0.019" ImageWireTM (LightLab Imaging, Westford, MA, USA) with automated pullback at 1 mm/s. Lumen measurements in both IVUS and OCT were performed with the same dedicated coronary vessel analysis software (CURAD, Wijk bij Duurstede, Netherlands).4-6
Histology Preparation and Analysis
Following the intracoronary imaging procedures, the specimens were sectioned for histological analysis. The distal side of the bifurcation of the LAD to the LCX was used as the starting point for the first histology section. From this point to the distal and proximal side, histology sections were taken at 5-mm intervals to the end of the specimen.7
Matching Procedure
The OCT and IVUS pullbacks were matched by vessel analysis software (CURAD), which can synchronize image data acquired with different modalities onto a single computer screen. The proximal and distal sheaths were used as longitudinal landmarks. After histology tissue processing, the sections (eg, microscopic images) were used to identify the corresponding location in the OCT image data set based on the similar visual appearance of the lumen morphology. Following this identification process, the visualization software synchronized the OCT and the IVUS image data sets and identified the corresponding IVUS image8 (Figure 1). To avoid the Dotter effect, only cross-sections where the IVUS catheter was moving freely in the lumen were selected.
Figure 1.Matching of IVUS, OCT, and histology ex vivo. The distal side of the bifurcation of the LAD to the LCX (*) was used as the starting point for the first histology section. From this point to the distal and proximal side, histology sections were taken at 5-mm intervals to the end of the specimen (A). The histology sections (B) were used to identify the corresponding location in the OCT image data set (C) based on the similar visual appearance of the lumen morphology. Following this identification, the visualization software synchronized the OCT and IVUS image data sets and identified the corresponding IVUS image (D).
In Vivo Human Coronary Arteries
Study Population
Patients who were undergoing OCT examination of 1 coronary artery were included in the study. The medical ethics committee of the Erasmus MC approved the study and all patients gave written informed consent.
IVUS Acquisition and Analysis
IVUS was acquired after intracoronary nitroglycerine administration using the Eagle Eye 20 MHz catheter (Volcano, Rancho Cordova, USA) with automatic continuous pullback at 0.5 mm/s. Before analysis, the IVUS data were retrospectively processed with an image-based gating system (Intelligate).9 The lumen was measured with the dedicated quantitative analysis software (CURAD).4,5
OCT Acquisition and Analysis
Optical coherence tomography acquisition was performed using the same system as described for the ex vivo study. Firstly, pullback was performed with proximal balloon occlusion and distal flush delivery, after which pullback was performed in the same segment without occlusion.
Occlusive technique: the occlusion balloon (Helios, Goodman, Japan) was advanced distal to the region of interest over a conventional angioplasty guidewire (0.014''). The conventional guidewire was then replaced by the OCT ImageWire and the occlusion balloon catheter was positioned proximal to the segment of interest. Pullback of the ImageWire was performed during inflation of the proximal occlusion balloon catheter at low pressure (0.4 atm) with simultaneous distal flush delivery (lactated Ringer's solution at 37oC; flow rate 0.8 mL/s). Images were acquired during pullback at 1 mm/s.
Non-occlusive technique: the ImageWire was placed in the artery using a double lumen catheter (Twin Pass catheter, Vascular Solutions). Pullback was performed during continuous injection of contrast medium (3 mL/s, Iodixanol 370, Visipaque, GE Health Care, Cork, Ireland) through the guide catheter with an injection pump. In this case, the automated pullback rate was 3 mm/s.
The lumen measurements were performed with proprietary software for offline analysis (LightLab Imaging, Westford, MA, USA). For all OCT studies, the settings for the refraction index were adjusted to the flush solution used (saline, Ringer's lactate, and x-ray contrast, respectively).
Matching of OCT and IVUS Pullbacks
The region of interest was matched in the 3 pullbacks (occlusive OCT, non-occlusive OCT, and IVUS) using landmarks such as side branches (Figure 2). Differences in pullback speed were adjusted to appropriately compare the 2 OCT pullbacks per frame.
Figure. 2.Selection of the region of interest in OCT pullbacks acquired with and without occlusion. PB1 and PB2 show the longitudinal view of the pullbacks acquired with and without occlusion, respectively. The white arrows indicate the side branches (SBA and SBB) used for matching of the region of interest. 1A and 2A show corresponding cross-sections of SBA with and without occlusion. 1B and 2B show corresponding cross-sections of the SBB with and without occlusion.
Statistical Analysis
Statistical analysis was performed using SPSS 12.0.1 for Windows (SPSS, Chicago, IL, USA). Continuous variables are expressed as mean (standard deviation) and categorical variables are expressed as percentages. The absolute and relative differences between measurements obtained with the different techniques were calculated. The relative difference was defined as the absolute difference divided by the average. Data are also expressed in Bland-Altman plots showing the absolute difference between corresponding lumen measurements for both techniques (y-axis) versus the average of both techniques (x-axis). The limits of agreement were calculated as the mean difference (+2SD).
RESULTS
Ex Vivo Human Coronary Arteries
Intravascular ultrasound and OCT were successfully performed in 8 out of 10 specimens. Intracoronary imaging was not possible in 2 cases due to the lumen being totally occluded. A good match between OCT, IVUS, and histology images was obtained in 35 frames. Overall, the mean lumen, plaque and vessel area when measured by histomorphometry were 2.5 (1.7) mm2, 3.7 (1.9) mm2, and 7.6 (3.2) mm2, respectively. The mean percentage of stenosis was 59% (18%).
Bland-Altman and regression analyses were performed on the lumen area measurements between the 3 imaging modalities (Figure 3). The average relative differences of the lumen measurements were 28% between OCT and histology, 40% between IVUS and histology, and 11% between IVUS and OCT.
Figure 3. Lumen measurements in ex vivo coronary arteries. Bland-Altman plots showing the differences in lumen measurements between OCT and histology (A), IVUS and histology (B), and OCT and IVUS (C). Correlation for lumen measurements between OCT and histology (A'), IVUS and histology (B'), and OCT and IVUS (C').
In Vivo Human Coronary Arteries
Clinical and Procedural Characteristics
Five patients (5 vessels) were included in the study. The average age was 61.2 (8.9) years and 4/5 were male. Three out of 5 had hypertension, 1 of 5 had diabetes, 4 of 5 had hyperlipidemia, and none were smokers. The imaged vessel was the LAD in 1 case and the LCX in the other 4 cases. Only 1 patient presented chest pain and ST depression during acquisition with balloon occlusion, which disappeared immediately after balloon deflation.
Comparison Lumen Measurements With OCT and IVUS
Overall, the mean lumen, plaque, and vessel area measured by IVUS were 6.3 (1.0) mm2, 6.9 (1.4) mm2, and 13.3 (1.7) mm2, and mean cross-sectional area stenosis was 52% (6%).
Comparison IVUS and OCT With Occlusion
The mean relative difference for the mean lumen area was 33.7%, and the mean absolute difference for the mean lumen area was 1.67 (0.54) mm2, with limits of agreement of 0.62-2.73 mm2 (Table 1). The mean relative difference for the minimum lumen area (MLA) between IVUS and OCT with occlusion was 55.5%, and the mean absolute difference was 1.90 (0.37) mm2, with limits of agreement of 1.17-2.63 mm2.
Comparison IVUS and OCT Without Occlusion
The average relative difference for the mean lumen area was 21.5%, and the absolute difference was 1.11 (0.53) mm2, with limits of agreement of 0.05-2.14 mm2 (Table 2). The average relative difference for MLA was 29%, and the average absolute difference was 1.10 (0.37) mm2, with limits of agreement 0.37-1.83 mm2.
Comparison of the Lumen Measurements by OCT With and Without Occlusion
For the comparison of lumen dimensions when measured by OCT, 373 matched frames obtained with and without occlusion were analyzed (Figure 4). The mean length of the region analyzed was 11.34 (3.87) mm with occlusion and 11.23 (3.86) mm without occlusion (P=.96). Bland-Altman and regression analyses were also performed for these measurements (Figure 5).
Figure 4.Example of differences in lumen measurements between OCT with and without occlusion. The figure shows corresponding images acquired with occlusion (A, B, and C) and without occlusion (A', B', and C'). The white arrows indicate the landmarks used for matching of the pullbacks (side branches in A and C and calcium spot in B). In all the examples, the lumen dimensions are smaller in the pullback acquired with occlusion.
Figure 5. Comparison of lumen measurements with optical coherence tomography (OCT) with and without occlusion. Bland-Altman plot showing the differences between lumen measurements in OCT pullbacks acquired with the occlusive and nonocclusive technique (A). Correlation between the lumen measurements in the OCT pullbacks acquired with and without occlusion (B).
The mean absolute and relative differences for mean lumen area were 0.61 (0.23) mm2 (13%) (limits of agreement, 0.15-1.07 mm2). The mean absolute and relative differences for MLA were 0.80 (0.21) mm2 (28%) (limits of agreement, 0.37-1.23 mm2) (Tables 3 and 4).
The mean lumen volumes were 55 (29.5) and 60.4 (30.4) mL3 with and without occlusion, respectively, with a relative difference of 11.7%. The mean absolute difference was 5.4 (2.7) mL3 and the limits of agreement were 0.0-10.7 mm3.
DISCUSSION
Optical coherence tomography is emerging as one of the most promising intracoronary imaging modalities due to its capacity to provide very high-resolution images of the coronary vessel wall. It enables the very detailed assessment of atherosclerotic plaque and the interaction between implanted coronary stents and the vessel wall.10,11 In recent years, there has been continuous technical development that has led to changes in the OCT acquisition technique (changing from the need to occlude the vessel to a non-occlusive procedure) which have simplified the use of the technology and reduced patient discomfort.12,13 However, even though the non-occlusive technique is currently gradually replacing the occlusive technique in many centers, it remains "off-label." Furthermore, many of the recently published OCT data were obtained by using the occlusive technique.14-20 The impact of the acquisition technique in relation to quantitative lumen measurements obtained by OCT and compared to quantitative IVUS has not yet been reported. The present study showed the following: a) in fixed human coronary arteries, both OCT and IVUS overestimated the lumen area compared to histology; b) lumen dimensions when measured by IVUS are larger than those when measured by OCT with or without occlusion in vivo; and c) the OCT acquisition technique (occlusive or non-occlusive) has an impact on lumen dimension measurements.
Differences in Lumen Measurements Between IVUS, OCT, and Histology
In the present study, the lumen areas when measured by IVUS and OCT were larger than those measured by histology in fixed human coronary arteries. The mean differences with histomorphometry were 0.8 mm2 (28%) for OCT and 1.3 mm2 (40%) for IVUS. These results are in line with a previous study in stented porcine coronary arteries that showed that lumen area was largest when measured by IVUS, followed by OCT, and was smallest when measured by histology.21 However, it should be borne in mind that the ex vivo processing of the specimen for histology analysis may have had an influence on lumen size. It is well established that tissue shrinkage occurs in different phases when preparing specimens for histology.22
Regarding the comparison of IVUS and OCT, our in vivo data are in agreement with the ex vivo findings. In our study, the lumen dimensions measured by IVUS in living patients were always higher than those measured by OCT with or without balloon occlusion. There are several studies in the literature comparing lumen dimensions measured by IVUS and OCT with occlusion that show contrasting results. Kawase et al, performed IVUS and OCT with occlusion in vivo in 6 coronary arteries in pigs and reported no differences in lumen areas and volumes between IVUS and OCT with occlusion.23 However, another study in pigs with a larger sample size reported an absolute difference of 0.49 mm2 between mean lumen area by IVUS and OCT acquired with the occlusive technique in stented segments.21 Yamaguchi et al evaluated in vivo differences in humans and found that the MLA was significantly smaller when measured by OCT with occlusion than when measured with IVUS (mean difference, 0.4 mm2). The greater differences found in our data could be explained by the fact that we did not include coronary segments treated with bare-metal stents. Bare-metal stents are probably less affected by variations in intracoronary pressure caused by proximal vessel occlusion as they have considerable radial strength.
To our knowledge, no published studies have reported differences in measurements between IVUS and OCT acquired with the non-occlusive technique. In the present study, the lumen dimensions were even larger by IVUS, but the differences, as expected, were smaller than with occlusion.
The greater overestimation of lumen dimensions when measured by IVUS compared to OCT could be related to the difficulty in differentiating the lumen border by IVUS due to blood speckle or presence of artefacts.24 OCT shows the lumen-intima interface as a sharply defined border, clearly visible in most cases.25 IVUS also overestimates the lumen area compared to quantitative angiography measurements (both by videodensitometry and edge detection).26-28 In our study, the differences between lumen measurements by IVUS and OCT were higher in vivo than ex vivo. This could be related to the frame selection for in vivo analysis. The IVUS pullback obtained in vivo was processed with a gating system that selects enddiastolic frames (the moment in the cardiac cycle when the vessel is biggest), whereas OCT pullback is not gated and therefore the frames could correspond to any moment in the cardiac cycle. Previous IVUS studies have suggested a variation in lumen area during the cardiac cycle in the range of 8%.29
The differences found between lumen size when measured by IVUS and OCT may have clinical implications if OCT is going to be used to define lesion severity. An MLA <4 mm2 has been associated with myocardial ischemia and has been considered a criterion for coronary lesion revascularization. In the present study, whereas all the patients had an MLA >4 mm2 when measured by IVUS, 4 out of 5 patients had an MLA <4 mm2 when measured by OCT with occlusion. All the patients were asymptomatic and were not treated. This reinforces the idea that IVUS criteria cannot be directly translated to OCT and emphasizes the need for new specific cut-off points to define lesion severity by OCT.
Impact of OCT Acquisition Technique on Lumen Measurements
The present report consistently shows that the mean and minimum lumen areas were smaller when OCT was performed with balloon occlusion than acquisitions without occlusion. These differences could be related to the decrease of intracoronary pressure due to the vessel occlusion that is not completely compensated by continuous flush injection. Furthermore, the injection of contrast medium during the non-occlusive procedure increases intracoronary pressure (Figure 6). These phenomena are likely to be more accentuated in healthy arteries and in coronary segments not treated with bare-metal stents (as in this study). The relative differences were higher in smaller vessels, suggesting that they may have a greater tendency to collapse after occlusion.
Figure 6.Intracoronary pressure recorded during saline or contrast injection. The figure shows the ECG (red) and intracoronary pressure (blue) records obtained in a left anterior descending coronary artery during injection of saline or contrast medium (Iodixanol 370). The distal intracoronary pressure was registered using the RADI pressure wire. The saline injection at 3 mL/s did not significantly modify intracoronary pressure (A). However, an increase in intracoronary pressure can be observed during contrast injection at 3 mL/s, (B). *Ventricular extrasystole.
Limitations
Even though a complex matching process was used, the difference in slice thickness between IVUS, OCT, and histology cross-sections may have influenced the results of the ex vivo study. The main limitations of the in vivo study are the small sample size and the limited type of coronary segments that could be analyzed. More data are needed to evaluate the quantitative differences between OCT pullbacks acquired with different techniques and IVUS in human coronary arteries treated with bare-metal stents. Further studies with a larger sample size and using different types of coronary segments are warranted before definite conclusions can be drawn.
CONCLUSION
In fixed human coronary arteries, both OCT and IVUS overestimated lumen areas compared to histomorphometry. Lumen dimensions measured by IVUS are larger than those measured by OCT with or without occlusion in vivo. The OCT acquisition technique (occlusive or non-occlusive) has an impact on quantitative lumen dimension measurements.
ABBREVIATIONS
IVUS: intravascular ultrasound
LAD: left anterior descending coronary artery
LCX: left circumflex artery
MLA: minimum lumen area
OCT: optical coherence tomography
SEE EDITORIAL ON PAGES 599-602
Correspondence: Dr E. Regar,
Thoraxcenter, Bd 585. 's-Gravendijkwal 230, 3015-CE Rotterdam, The Netherlands
E-mail: e.regar@erasmusmc.nl
Received January 13, 2009.
Accepted for publication February 19, 2009.