Publish in this journal
Journal Information
Vol. 72. Issue 1.
Pages 78 (January 2019)
Share
Share
Download PDF
More article options
Vol. 72. Issue 1.
Pages 78 (January 2019)
ECG Contest
DOI: 10.1016/j.rec.2018.05.005
Full text access
Response to ECG, December 2018
Respuesta al ECG de diciembre de 2018
Visits
...
Ana Fernández-Vegaa,
Corresponding author
afvega@salud.madrid.org

Corresponding author:
, Daniel Enríquez-Vázqueza, Julián Palacios-Rubiob
a Instituto Cardiovascular, Hospital Clínico San Carlos, Madrid, Spain
b Unidad de Arritmias, Instituto Cardiovascular, Hospital Clínico San Carlos y CIBERCV, Madrid, Spain
Related content
Rev Esp Cardiol. 2018;71:107210.1016/j.rec.2018.05.006
Ana Fernández-Vega, Daniel Enríquez-Vázquez, Julián Palacios-Rubio
This item has received
...
Visits
(Daily data update)
Article information
Full Text
Bibliography
Download PDF
Statistics
Figures (1)
Full Text

The correct answer is 3. There is no sensing malfunction (answer 2, incorrect): if the cardiac resynchronization therapy pacemaker (CRT-P) had not detected the ventricular extrasystole, this pacing pulse would have occurred as for the others at 840ms; however, it comes earlier, at 700ms. To maintain optimal biventricular pacing, CRT-Ps usually have a function denoted ventricular sense response. In the event of a conducted beat or extrasystole sensed in the right ventricle (RV), a pacing pulse is emitted.1 The CRT-P telemetry (Figure) shows resynchronized beats (BV) with a sensed beat (SV), superimposed with resynchronization pacing. In the second channel of the Figure, corresponding to lead I of the ECG, 6 resynchronized beats are shown (biventricular [BV] pacing), whereas the seventh (Figure, asterisk) is an extrasystole (sensed ventricle [SV]) which the CRT-P paces automatically to attempt resynchronization (superimposed BV). The right bundle branch block morphology of the ventricular extrasystoles in the ECG indicates that they originate from the left ventricle. The ventricular sense response function uses RV sensing, and so there is a delay between onset of QRS in the left ventricle, RV sensing, and subsequent pacing: in the third channel, which shows the local RV electrogram, the extrasystole occurs later than the initial QRS in the surface ECG.

(0.15MB).

The first channel of the Figure shows the pacing channel and the atrial electrogram but, given that the atrial port is plugged, far-field activity can be observed.

It is likely that the CRT-P needs to be optimized (response 1 incorrect and response 3 correct) as baseline QRS is negative in V1 and positive in D-I.2 Finally, when the elective replacement indicator of the CRT-P is triggered, biventricular pacing is maintained at the lower rate limit,1 which would not explain the early spikes (response 4, incorrect).

References
[1]
Guía de Referencia de los Marcapasos Medtronic Consulta CRT-P. Minneapolis: Medtronic; 2010. Available from: http://manuals.medtronic.com/
[2]
V. Ganière, G. Domenichini, V. Niculescu, R. Cassagneau, P. Defaye, H. Burri.
A new electrocardiogram algorithm for diagnosing loss of ventricular capture during cardiac resynchronisation therapy.
Europace., 15 (2013), pp. 376-381
Copyright © 2018. Sociedad Española de Cardiología
Idiomas
Revista Española de Cardiología (English Edition)

Subscribe to our newsletter

Article options
Tools
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

es en
Política de cookies Cookies policy
Utilizamos cookies propias y de terceros para mejorar nuestros servicios y mostrarle publicidad relacionada con sus preferencias mediante el análisis de sus hábitos de navegación. Si continua navegando, consideramos que acepta su uso. Puede cambiar la configuración u obtener más información aquí. To improve our services and products, we use "cookies" (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.