Transfemoral Rescue for a Transapical Malpositioned TAVI Device
Rescate transfemoral en un dispositivo TAVI transapical mal posicionado
Visits
5665
Vincent J. Nijenhuis
Corresponding author
, Martin J. Swaans, Jurrien M. Ten Berg
Department of Cardiology, St. Antonius Hospital, Nieuwegein, Netherlands
This item has received
Article information
Full Text
Download PDF
Statistics
Figures (3)
Show moreShow less
Full Text

An 81 year-old male was referred for transcatheter aortic valve implantation (TAVI) because of a degenerated Carpentier Edwards Perimount 25mm (2005). He had a coronary artery bypass graft, involving a left internal mammary artery to the left anterior descending coronary artery and a saphenous vene to the posterior descending artery. Multidetector row computed tomography imaging showed iliac arteries with a lumen of 5.5-6.0mm and a tortuous course with severe calcifications (Figures 1 and 2). Due to the risk of vascular complications associated with a transfemoral access, the patient was accepted for a transapical TAVI using the JenaValve 23mm (JenaValve Technology GmbH, Germany).

Figure 1
(0.07MB).
Figure 2
(0.07MB).

During the procedure, we opened the sheath releasing the 3 “positioning feelers” of the JenaValve. Upon positioning, the clips were released impeding fixation of the device to the bioprosthesis (Figure 3A). We released the JenaValve in the ascending aorta where it remained rocking, potentially blocking the ostium of the brachiocephalic trunc. We performed a transfemoral rescue with a CoreValve bioprosthesis (Medtronic, United States). We advanced the CoreValve through the JenaValve without displacing it (Figure 3B). We then successfully deployed it in the bioprosthesis (Figure 3C).

Figure 3
(0.12MB).

After releasing the feelers, an automatic stop in the Cathlete Plus Delivery Systems prevents it from rotating further until the safety button is pushed. In this case, the clipping mechanism was prematurely released due to a faulty spring in the safety button mechanism. Because of similar cases, this lead to a temporary voluntarily recall of all Cathlete Plus Delivery Systems from the market in 2014.

Copyright © 2016. Sociedad Española de Cardiología
Idiomas
Revista Española de Cardiología (English Edition)

Subscribe to our newsletter

View newsletter history
Article options
Tools
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?