Two Different Sites of Membrane Rupture in the Berlin-Heart EXCOR Ventricular Assist Device
Dos lugares diferentes de rotura de membrana en la asistencia Berlin-Heart Excor
Visits
...
Iago Sousa Casasnovasa,
Corresponding author
iagosousa@yahoo.es

Corresponding author:
, Pablo Díez Villanuevaa, Felipe Díez del Hoyoa, Manuel Ruiz Fernándezb, Ángel González Pintob, Francisco Fernández-Avilésa
a Servicio de Cardiología, Hospital General Universitario Gregorio Marañón, Madrid, Spain
b Servicio de Cirugía Cardiaca, Hospital General Universitario Gregorio Marañón, Madrid, Spain
This item has received
...
Visits
(Daily data update)
Article information
Full Text
Bibliography
Download PDF
Statistics
Figures (2)
Additional material (2)
Full Text
To the Editor,

The Berlin-Heart EXCOR ventricular assist device (Berlin-Heart AG, Germany) is a paracorporeal pulsed flow device driven by a pneumatic mechanism. Currently, the main application is as a bridge to heart transplant, both in adults and children.1,2

The device comprises a pneumatic driving unit, a system of tubes, and a ventricle or pump. This pump is divided into 2 chambers (blood chamber and air chamber) by a 3-layer polyurethane membrane. The negative and positive pressure in diastole and systole, respectively, generated by the drive unit and transmitted to the ventricle, displace the membrane, with the corresponding effect of suction or blood expulsion (Figure 1). Mechanical problems occur more frequently in pulsed flow devices than in continuous flow ones.3 Membrane rupture is, however, an uncommon problem and has been reported in a limited number of cases,4–6 after more than 2000 implants up until 2015. We present 2 cases of membrane rupture in our center.

Figure 1.

Diagram of the functioning of the Berlin-Heart EXCOR ventricular assist device.

(0.14MB).

The first case occurred in a 62-year-old man, with a biventricular Berlin-Heart EXCOR implant after acute myocardial infarction complicated with cardiogenic shock as a bridge to heart transplant. On day 77 of treatment, while the patient was on the ward, a mobile shadow, measuring 3cm across, was detected in the right pump, with an irregular border and fine bands of adhered fibrin. This image was apparent in systole and disappeared in diastole (Figure 2A and video 1 of the supplementary material). It was only visible in the blood chamber; the membrane appeared normal from the air chamber. The patient was asymptomatic and hemodynamically stable, with no warning alarms showing on the display of the operating panel. Membrane rupture was suspected and an emergency pump change was undertaken. After explantation, a torn area was detected on the internal layer of the membrane (adjacent to the blood chamber).

Figure 2.

Diagram of membrane rupture. A: Rupture of the internal layer (case 1). B: Rupture of the external layer (case 2).

(0.22MB).

The second case occurred in a 63-year-old man with dilated cardiomyopathy, who had been implanted with a Berlin-Heart EXCOR device as a bridge to heart transplant. On day 161 of assist, while on the ward, the patient suddenly experienced dyspnea, malaise, and cold sweat. The physical examination confirmed cardiogenic shock, with a warning of “low left ventricular flow” showing on the display of the operating panel. A perfectly round shadow measuring 5cm in diameter was detected in the blood chamber. This shadow was more evident during device systole (Figure 2B and video 2 of the supplementary material). In the air chamber, a rounded bulge was observed in systole, while the appearance was normal during diastole. Both the shadow and bulge became increasingly marked in the course of minutes. Membrane rupture was suspected and an emergency left pump change was undertaken. After explantation, a small tear (measuring a few millimeters) could be seen on the outer layer of the membrane (adjacent to the air chamber) and air was trapped between the external and middle layer.

Both cases illustrate 2 different presentations of membrane rupture. To date, cases of rupture of the internal and external layer of the membrane have been described (with no reports of rupture of the middle layer).

Rupture of the internal layer (first patient) meant that the blood came into contact with the nonbiocompatible surfaces between the internal and middle layer, thereby favoring the formation of fibrin and thrombi in this area. The middle and external layers remained intact and pump function was maintained. As the tear spread, a flap could be detected in the blood chamber. This appeared and disappeared during the device cycle, with an irregular border and chaotic movement (Figure 2A). By contrast, the appearance from the air chamber was normal. This type of rupture, therefore, increases the risk of thrombosis, and it is hemodynamically tolerated because the assist device continues to function correctly.

Rupture of the outermost layer (second case) led to progressive air entrapment between the middle and external layer: the air entered in the space during systole, and its exit in diastole was hindered by the valve effect generated by the suction of the device (Figure 2B). External examination revealed a rounded shadow in the blood chamber during systole, whereas a rounded bulge was seen in the air chamber during systole. This mechanism of air entrapment led to progressive pump dysfunction (pump obstruction) that had a substantial hemodynamic impact on the patient. In this situation, cardiogenic shock can develop in a question of minutes.

The 2 cases presented and the schematics help us to understand the development of this complication, each one with a different appearance on examination, different clinical course, and different associated complications. In our opinion, awareness of this complication and its pathophysiology is important for early detection, as emergency replacement of the ventricle is essential to avoid adverse outcomes.

Material adicional

Video 1 Rupture of the internal layer.

Video 2 Rupture of the external layer.

(3.32MB)
(0.89MB)

References
[1]
J.K. Kirklin, D.C. Naftel, F.D. Pagani, R.L. Kormos, L.W. Stevenson, E.D. Blume, et al.
Sixth INTERMACS annual report: a 10,000-patient database.
J Heart Lung Transplant., 33 (2014), pp. 555-564
[2]
C. Schmid, T. Tjan, C. Etz, H. Welp, A. Rukosujew, S. Klotz, et al.
The Excor device, revival of an old system with excellent results.
Thorac Cardiovasc Surg., 54 (2006), pp. 393-399
[3]
M.S. Slaughter, J.G. Rogers, C.A. Milano, S.D. Russell, J.V. Conte, D. Feldman, et al.
Advanced heart failure treated with continuous-flow left ventricular assist device.
N Engl J Med., 361 (2009), pp. 2241-2251
[4]
F.J. Hernández-Pérez, R.J. Burgos-Lázaro, M.F. Gómez-Bueno.
Una complicación grave en paciente con asistencia ventricular.
Rev Esp Cardiol., 67 (2014), pp. 322
[5]
S. Volz, M. Holmberg, B. Redfors, G. Dellgren.
Acute tamponade of the left paracorporeal pump house due to membrane defect in a patient with a Berlin Heart EXCOR biventricular assist device.
Eur J Cardiothorac Surg., 46 (2014), pp. 743-744
[6]
E. Pektok, S. Kucukaksu.
Pulsatile paracorporeal pump membrane rupture: clinical advantage of a multilayer membrane.
Copyright © 2016. Sociedad Española de Cardiología
Idiomas
Revista Española de Cardiología (English Edition)

Subscribe to our newsletter

Article options
Tools
Supplemental materials
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

es en
Política de cookies Cookies policy
Utilizamos cookies propias y de terceros para mejorar nuestros servicios y mostrarle publicidad relacionada con sus preferencias mediante el análisis de sus hábitos de navegación. Si continua navegando, consideramos que acepta su uso. Puede cambiar la configuración u obtener más información aquí. To improve our services and products, we use "cookies" (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.